Standard
Configuration
Maximum Operating Pressure: 10,000
psi. (68.9 MPa).
Pressure Display: LED.
± 1 psi. (6.9 kPa).
Temperature Range: Ambient
to 200˚C.
Temperature Precision: ±0.5˚C.
Temperature Display: LED.
Displays internal vessel temperature.
Flow Rates: Up to 250
grams/min. (330 ml/min.)
liquid CO2 under standard
operating conditions.
Restrictor Valve: Heated
up to 250˚C; resistant to blockage (factory set to 80°C).
Extraction/Reaction
Vessel Size: Up to 2 liters; vessels are user interchangeable.
Collection Vessel: Externally
mounted. Many options are available.
Heating Power: All
extraction/reaction vessels are heated with band heaters (2000 watts
max.).
Over-Pressure Safeguards:
Rupture disc assembly.
|
Instrument Control: Vessel
and preheater temperature controlled by PID-Fuzzy Logic Controllers.
Pressure controlled by manual regulator.
Dimensions: Width: 61
cm, Depth: 61 cm, Height: 96 cm.
Weight (excluding
vessel): 60 kg (130 Lb).
Configuration
Options
Custom Extraction and
Reaction Vessels: Multiple vessels, stirring, windows and a selection
of vessel sizes are available.
Co- solvent Addition
Module: Manual doping or direct, in-line metered addition.
Pre heater: Improves
temperature consistency of the fluid in large vessels by heating the fluid
before it reaches the main pressure vessel. Recommended for all extraction
systems.
System
Requirements
Power Requirements (U.S.):
110 VAC., single phase, 15 amps (International power available).
Gas Supply: Liquid CO2
cylinder with dip tube.
House Air: Dry air,
regulated to 110 psi. (760 kPa). |